Remote Control
R&S
®
ESRP
813
User Manual 1175.7445.02 ─ 08
Suffix:
<n>
.
window; For applications that do not have more than 1 measure-
ment window, the suffix <n> is irrelevant.
<t>
irrelevant
Parameters:
<Mode>
ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST:
ABS
Example:
DISP:TRAC:Y:MODE REL
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
<ReferenceLevel>
This command defines the reference level.
With the reference level offset
≠
0, the value range of the reference level is modified by
the offset.
The unit depends on the setting defined with
Suffix:
<n>
.
irrelevant.
<t>
irrelevant
Parameters:
<ReferenceLevel>
The unit is variable.
Range:
see datasheet
*RST:
-10dBm
Example:
DISP:TRAC:Y:RLEV -60dBm
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
<Value>
This command defines a reference level offset.
Suffix:
<n>
.
irrelevant.
<t>
irrelevant
Parameters:
<Value>
Range:
-200 to 200
*RST:
0
Default unit: dB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Remote Commands in Spectrum Analyzer Mode