Remote Control
R&S
®
ESRP
657
User Manual 1175.7445.02 ─ 08
CALCulate<n>:MARKer<m>:FUNCtion:CSTep
This command matches the frequency step size to the current marker frequency.
The command turns delta markers into normal markers.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Example:
CALC:MARK3:FUNC:CST
Sets the center frequency to the same value as the frequency of
marker 3.
Usage:
Event
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:ZOOM
<Range>
This command defines the range to be zoomed around marker 1. Marker 1 is activated
first, if necessary.
The marker frequency becomes the new receiver or center frequency and the span is
adjusted according to the zoom factor.
Note that you should perform a complete measurement with synchronization to the end
of the measurement. This is only possible for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<Range>
<numeric_value>
Example:
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:ZOOM 1kHz;*WAI
Activates zooming and waits for its end.
Manual operation:
See
CALCulate<n>:MARKer<m>:SCOupled[:STATe]
<State>
This command couples or decouples the marker frequency to the scan range settings.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
CALC:MARK:SCO ON
Couples the scan range settings to the marker frequency.
Manual operation:
See
Remote Commands in Receiver Mode