Remote Control
R&S
®
ESRP
778
User Manual 1175.7445.02 ─ 08
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:SDEV ON
Switches on the function.
CALC:MARK:FUNC:SUMM:PHOL ON
Switches on the peak value measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:SDEV:PHOL:RES?
Outputs the result.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:RESult?
This command queries the standard deviation of the time domain power.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Return values:
<StandardDeviation> Standard deviation of the signal during the measurement time.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:SDEV ON
Switches on the function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:SDEV:RES?
Outputs the result.
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation[:STATe]
<State>
This command turns the measurement of the standard deviation of the time domain
power on and off.
Note
: The measurement is performed on the trace marker 1 is positioned. In order to
evaluate another trace, marker 1 must be positioned on another trace with
.
Suffix:
<n>
.
Selects the measurement window.
Remote Commands in Spectrum Analyzer Mode