Remote Control
R&S
®
ESRP
723
User Manual 1175.7445.02 ─ 08
Suffix:
<range>
.
1...20
range
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
LIST:RANG3:INP:GAIN:STATe ON
Switches the preamplifier for range 3 on or off.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:LIMit:STARt
<Limit>
This command sets an absolute limit value at the start frequency of the specified
range. Different from manual operation, this setting is independent of the defined limit
check type.
Suffix:
<range>
.
1...20
range
Parameters:
<Limit>
-400 to in 400 dBm
*RST:
13 dBm
Example:
LIST:RANG1:LIM:ABS:STAR 10
Sets an absolute limit of 10 dBm at the start frequency of the
range.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:LIMit:STATe
This command activates or deactivates the limit check for all ranges.
Suffix:
<range>
.
irrelevant
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
LIST:RANG3:LIM:STAT ON
Activates a limit check for all ranges.
Manual operation:
See
Remote Commands in Spectrum Analyzer Mode