Remote Control
R&S
®
ESRP
774
User Manual 1175.7445.02 ─ 08
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Return values:
<PeakPower>
Peak power of the signal during the measurement time.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:PPE ON
Switches on the function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:PPE:RES?
Outputs the result.
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak[:STATe]
<State>
This command turns the measurement of the positive peak time domain power on and
off.
Note
: The measurement is performed on the trace marker 1 is positioned. In order to
evaluate another trace, marker 1 must be positioned on another trace with
.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK:FUNC:SUMM:PPE ON
Switches on the function.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:AVERage:RESult?
This command queries the average RMS time domain power. The query is only possi-
ble if averaging has been activated previously using
.
Remote Commands in Spectrum Analyzer Mode