Remote Control
R&S
®
ESRP
858
User Manual 1175.7445.02 ─ 08
TRACe<n>:IQ:RLENgth
<NoOfSamples>
This command sets the record length for the acquired I/Q data. Increasing the record
length automatically also increases the measurement time.
Note
: Alternatively, you can define the measurement time using the
SENS:SWE:TIME
command.
Suffix:
<n>
.
irrelevant
Parameters:
<NoOfSamples>
Number of measurement values to record.
Range:
1 ... 209715200(= 200*1024*1024);
*RST:
691
Example:
TRAC:IQ:RLEN 256
Manual operation:
See
TRACe<n>:IQ:SET
NORM,<Placeholder>, <SampleRate>, <TriggerMode>,
<TriggerSlope>, <PretriggerSamp>, <NumberSamples>
This command defines the settings of the R&S
ESRP hardware for the measurement
of I/Q data. This allows setting the sample rate, trigger conditions and the record
length.
If this command is omitted, the current instrument settings are used for the correspond-
ing parameters.
This command switches to IQ mode automatically (see also
The trigger level can be set using the
command.
Note:
When using the default settings with the
following minimum buffer sizes for the response data are recommended: ASCII format
10 kBytes, Binary format 2 kBytes.
Tip:
To determine the trigger point in the sample (for "External" or "IF Power" trigger
Parameters:
NORM
is not evaluated, but must be inserted
<Placeholder>
numeric value
is not evaluated, but must be inserted
<TriggerMode>
IMMediate | EXTernal | IFPower | BBPower | PSE
Selection of the trigger source used for the measurement.
For IMM mode, gating is automatically deactivated.
*RST:
IMM
Remote Commands in I/Q Analyzer Mode