System Configuration
R&S
®
ESRP
486
User Manual 1175.7445.02 ─ 08
Furthermore, some special GPIB settings are available to emulate HP models (see
Chapter 11.18, "GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and
8594E"
●
●
●
●
"Default"
Standard R&S
"R&S FSP"
Compatible to R&S FSP, see
"R&S FSU"
Remote command:
Mode Default ← Compatibility Mode ← GPIB ← General Setup
Resets the number of measurement points and available bandwidths to default
R&S
ESRP values.
Remote command:
SYST:COMP DEF
, see
Mode R&S FSP ← Compatibility Mode ← GPIB ← General Setup
Sets the number of measurement points and available bandwidths as in R&S FSP
devices.
Remote command:
SYST:COMP FSP
, see
Mode R&S FSU ← Compatibility Mode ← GPIB ← General Setup
Sets the number of measurement points and available bandwidths as in R&S FSU
devices.
Remote command:
SYST:COMP FSU
, see
GPIB Language ← Compatibility Mode ← GPIB ← General Setup
Opens a list of selectable remote-control languages:
Language
Comment
SCPI
71100C
Compatible to 8566A/B
71200C
Compatible to 8566A/B
71209A
Compatible to 8566A/B
8560E
8561E
8562E
8563E
Instrument Setup and Interface Configuration – SETUP Key