LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
701
User Manual 1173.9628.02 ─ 16
Example:
See
Usage:
Query only
Firmware/Software:
V2.0.20
Options:
R&S CMW-KM012
FETCh:LTE:MEAS<i>:MEValuation:LIST:SEGMent<no>:PMONitor:ARRay:
STARt?
Returns the offset of the first power monitor result for segment <no> within a result list
for all measured segments. Such a result list is e.g. returned by the command
.
A returned <Start> value n indicates that the result for the first subframe of the seg-
ment is the (n+1)
th
result in the power result list over all segments.
Suffix:
<no>
.
1..1000
Return values:
<Reliability>
In list mode, a zero reliability indicator indicates that the results
in all measured segments are valid. A non-zero value indicates
that an error occurred in at least one of the measured segments.
<Start>
Range:
0 to 3999
Example:
See
Usage:
Query only
Firmware/Software:
V2.0.20
V2.1.25: increased maximum number of segments to 250
V2.1.30: increased maximum number of segments to 512
V3.0.50: increased maximum number of segments to 1000
V3.2.70: increased maximum start value to 3999
Options:
R&S CMW-KM012
FETCh:LTE:MEAS<i>:MEValuation:LIST:SEGMent<no>:PMONitor:ARRay:
LENGth?
Returns the number of power monitor results for segment <no> contained in a result
list for all measured segments. Such a result list is e.g. returned by the command
.
Suffix:
<no>
.
1..1000
Command Reference