LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
611
User Manual 1173.9628.02 ─ 16
CONFigure:LTE:MEAS<i>:MEValuation:SCOunt:MODulation
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<StatisticCount>
Number of measurement intervals (slots)
Range:
1 slot to 1000 slots
*RST:
20 slots
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.10.1
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:SRS:ENABle
<Enable>
Specifies whether the uplink signal may contain a sounding reference signal (ON) or
not (OFF). As a consequence, the last SC-FDMA symbol of each subframe is ignored
or evaluated for modulation results.
For the combined signal path scenario, use
.
Parameters:
<Enable>
OFF | ON
OFF
: all SC-FDMA symbols are evaluated
ON
: the last SC-FDMA symbol of each subframe is not evalu-
ated
*RST:
OFF
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V3.0.10
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:MODulation:EWLength
<CPnormB014>,
<CPnormB030>, <CPnormB050>, <CPnormB100>, <CPnormB150>,
<CPnormB200>, <CPextB014>, <CPextB030>, <CPextB050>, <CPextB100>,
<CPextB150>, <CPextB200>
Specifies the EVM window length in samples for all channel bandwidths, depending on
the cyclic prefix (CP) type.
Command Reference