LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
122
User Manual 1173.9628.02 ─ 16
Fig. 2-47: I/Q settings (1 Cell - Fading - 1 RF Out, external fading)
................................................................................................ 122
.............................................................................................122
Connector (Out / In)
Select the output connector. The input connector depends on the output connector and
is displayed for information.
The DIG IQ connectors are located at the rear panel (if an I/Q board is installed).
Remote command:
Sample Rate (Out / In)
The used sample rate is displayed for information. The value is fixed.
Configure the connected instrument accordingly (baseband input settings and digital
I/Q output settings).
Remote command:
Baseband PEP (Out / In)
Indicates the peak envelope power of the baseband signal as dB value relative to full
scale. "Full scale" in this case corresponds to the maximum representable amplitude of
the I/Q samples.
Use the displayed output PEP value to configure the baseband input of the connected
instrument.