LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
177
User Manual 1173.9628.02 ─ 16
Fig. 2-77: Connected DRX settings (values used in example)
The unit "PDCCH subframe" used in this context refers to subframes with PDCCH. For
FDD this is any subframe. For TDD only DL subframes and subframes with DwPTS
are counted.
...........................................................................................178
......................................................................................... 178
..............................................................................................179
Connected DRX
Enables or disables DRX and selects a set of DRX settings.
If CQI reporting is enabled, connected DRX is not possible (disabled and grayed out).
Option R&S CMW-KS510 is required.
"DRX_S"
Settings according to 3GPP TS 36.521-3, table H.3.6-1.
"DRX_L"
Settings according to 3GPP TS 36.521-3, table H.3.6-2.
"User-Defined" You can configure all DRX settings.
Remote command:
On Duration Timer
"On" time at the beginning of each short or long DRX cycle (onDurationTimer).
Remote command:
GUI Reference