LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
601
User Manual 1173.9628.02 ─ 16
CONFigure:LTE:MEAS<i>:MEValuation:PLCid
................................................................. 608
CONFigure:LTE:MEAS<i>:MEValuation:DSSPusch
...........................................................609
CONFigure:LTE:MEAS<i>:MEValuation:GHOPping
...........................................................609
CONFigure:LTE:MEAS<i>:MEValuation:MSUBframes
....................................................... 609
CONFigure:LTE:MEAS<i>:MEValuation:TOUT
<Timeout>
Defines a timeout for the measurement. The timer is started when the measurement is
initiated via a
READ
or
INIT
command. It is not started if the measurement is initiated
manually (ON/OFF key or RESTART/STOP key).
When the measurement has completed the first measurement cycle (first single shot),
the statistical depth is reached and the timer is reset.
If the first measurement cycle has not been completed when the timer expires, the
measurement is stopped. The measurement state changes to
RDY
and the reliability
indicator is set to 1, indicating that a measurement timeout occurred. Still running
READ
,
FETCh
or
CALCulate
commands are completed, returning the available results.
At least for some results there are no values at all or the statistical depth has not been
reached.
A timeout of 0 s corresponds to an infinite measurement timeout.
Parameters:
<Timeout>
Default unit: s
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V2.0.10
CONFigure:LTE:MEAS<i>:MEValuation:REPetition
<Repetition>
Specifies the repetition mode of the measurement. The repetition mode specifies
whether the measurement is stopped after a single-shot or repeated continuously. Use
CONFigure:..:MEAS<i>:...:SCOunt
to determine the number of measurement
intervals per single shot.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<Repetition>
SINGleshot | CONTinuous
SINGleshot:
Single-shot measurement
CONTinuous:
Continuous measurement
*RST:
SING
Example:
See
Single-Shot and Continuous Measurements
Firmware/Software:
V1.0.10.1
Manual operation:
See
Command Reference