LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
617
User Manual 1173.9628.02 ─ 16
CONFigure:LTE:MEAS<i>:MEValuation:PDYNamics:AEOPower:LAGGing
<Lagging>
Shifts the end of the evaluation period for OFF power measurements.
Parameters:
<Lagging>
Positive values reduce the evaluation period (ends earlier).
Negative values increase the evaluation period (ends later).
Range:
-1000 Ts to 1000 Ts
*RST:
0 Ts
Default unit: Ts
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V3.2.70
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:SCOunt:POWer
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<StatisticCount>
Number of measurement intervals
Range:
1 subframe to 1000 subframes
*RST:
20 subframes
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V2.0.20
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:POWer:HDMode
<HighDynamicMode>
Enables or disables the high dynamic mode for power dynamics measurements.
Parameters:
<HighDynamicMode> OFF | ON
*RST:
OFF
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V2.1.25
Manual operation:
See
Command Reference