![R&S CMW-KM5 Series Скачать руководство пользователя страница 265](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802265.webp)
LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
265
User Manual 1173.9628.02 ─ 16
SENSe:LTE:SIGN<i>:UEReport[:PCC]:RSRQ:RANGe?
SENSe:LTE:SIGN<i>:UEReport:SCC:RSRQ:RANGe?
Returns the RSRQ value range, corresponding to the RSRQ index reported by the UE
for PCC or SCC.
Return values:
<Lower>
Range:
-19.5 dB to -3 dB
Default unit: dB
<Upper>
Range:
-19.5 dB to -3 dB
Default unit: dB
Usage:
Query only
Firmware/Software:
V2.0.10
V3.2.70: added SCC command
Options:
SCC: R&S CMW-KS502/-KS552 for FDD/TDD
Manual operation:
See
SENSe:LTE:SIGN<i>:UEReport[:PCC]:SCELl?
SENSe:LTE:SIGN<i>:UEReport:SCC:SCELl?
Returns measurement report values for the PCC or SCC of the serving LTE cell.
Return values:
<RSRP>
RSRP as dimensionless index
Range:
0 to 97
<RSRQ>
RSRQ as dimensionless index
Range:
0 to 34
Example:
See
Querying UE Measurement Report Contents
Usage:
Query only
Firmware/Software:
V3.0.50
V3.2.70: added SCC command
Options:
SCC: R&S CMW-KS502/-KS552 for FDD/TDD
Manual operation:
See
SENSe:LTE:SIGN<i>:UEReport[:PCC]:SCELl:RANGe?
SENSe:LTE:SIGN<i>:UEReport:SCC:SCELl:RANGe?
Returns the value ranges corresponding to the dimensionless index values reported for
the PCC or SCC of the serving LTE cell.
Return values:
<RSRPlower>
RSRP minimum value
Range:
-140 dBm to -44 dBm
Default unit: dBm
Command Reference