LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
681
User Manual 1173.9628.02 ─ 16
<2_OutOfTol>
Out of tolerance result, i.e. percentage of measurement inter-
vals of the statistic count for spectrum emission measurements
exceeding the specified spectrum emission mask limits.
Range:
0 % to 100 %
Default unit: %
<3_CurrNeg1> ...
<14_CurrNeg12>
<15_CurrPos1> ...
<26_CurrPos12>
Margin results for current trace
Range:
-160 dB to 160 dB
Default unit: dB
<27_AvgNeg1> ...
<38_AvgNeg12>
<39_AvgPos1> ...
<50_AvgPos12>
Margin results for average trace
Range:
-160 dB to 160 dB
Default unit: dB
<51_MinNeg1> ...
<62_MinNeg12>
<63_MinPos1> ...
<74_MinPos12>
Margin results for maximum trace (resulting in minimum mar-
gins)
Range:
-160 dB to 160 dB
Default unit: dB
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.2.80
FETCh:LTE:MEAS<i>:MEValuation:SEMask:MARGin:CURRent:NEGativ?
FETCh:LTE:MEAS<i>:MEValuation:SEMask:MARGin:CURRent:POSitiv?
FETCh:LTE:MEAS<i>:MEValuation:SEMask:MARGin:AVERage:NEGativ?
FETCh:LTE:MEAS<i>:MEValuation:SEMask:MARGin:AVERage:POSitiv?
FETCh:LTE:MEAS<i>:MEValuation:SEMask:MARGin:MINimum:NEGativ?
FETCh:LTE:MEAS<i>:MEValuation:SEMask:MARGin:MINimum:POSitiv?
Returns spectrum emission mask margin results. A negative margin indicates that the
trace is located above the limit line, i.e. the limit is exceeded.
The individual commands provide results for the
CURRent
,
AVERage
and maximum
traces (resulting in
MINimum
margins). For each trace, the X and Y values of the mar-
gins for emission mask areas 1 to 12 are provided for
NEGative
and
POSitive
offset
frequencies. For inactive areas, NCAP is returned.
The number to the left of each result parameter is provided for easy identification of the
parameter position within the result array.
Return values:
<1_Reliability>
<2_OutOfTol>
Out of tolerance result, i.e. percentage of measurement inter-
vals of the statistic count for spectrum emission measurements
exceeding the specified spectrum emission mask limits.
Range:
0 % to 100 %
Default unit: %
Command Reference
深圳德标仪器
135-1095-0799