LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
548
User Manual 1173.9628.02 ─ 16
Stop Condition
Specifies the conditions for an early termination of the measurement:
●
None:
The measurement is performed according to its "Repetition" mode and "Sta-
tistic Count", irrespective of the limit check results.
●
On Limit Failure:
The measurement is stopped as soon as one of the limits is
exceeded, irrespective of the repetition mode set. If no limit failure occurs, it is per-
formed according to its "Repetition" mode and "Statistic Count". Use this setting for
measurements that are essentially intended for checking limits, e.g. production
tests.
Remote command:
Measurement Mode
Selects/displays the measurement mode.
●
Normal
:
In normal mode, the measurement uses a constant reference level for the entire
measurement duration.
Use the normal mode for the standalone scenario. For the other scenarios, the nor-
mal mode is only suitable, if the controlling application does not modify the refer-
ence level during the measurement.
●
TPC Mode
:
In TPC mode, the measurement (re)configures the reference level once per sub-
frame. With the combined signal path scenario, this is done according to the infor-
mation provided by the controlling application. With the standalone scenario, the
reference levels to be used must be configured, see parameter
.
The TPC mode has been designed for TPC measurements where the UE power is
ramped up or down over a wide range, exceeding the dynamic range of the mea-
surement. A typical example is a conformance test case according to 3GPP TS
36.521, section "6.3.5.2 Power Control Relative power tolerance".
Note that for spectrum measurements the normal mode is recommended. For
channel bandwidths > 10 MHz, spectrum measurements are not supported in TPC
mode.
●
Multi Eval List Mode
:
The list mode can only be enabled via the remote control command listed below.
When it has been enabled, "Multi Eval List Mode" is displayed as measurement
mode. You can disable it by selecting one of the other modes.
For an introduction to the list mode see
Option R&S CMW-KM012 is required.
While the combined signal path scenario is active, this parameter is not configurable.
Remote command:
TPC Mode
This node configures a sequence of expected nominal power values to be applied dur-
ing a measurement in TPC mode, using the standalone scenario. It is only visible, if the
measurement mode "TPC Mode" is selected.
GUI Reference