LTE PRACH Measurement
R&S
®
CMW-KM5xx/-KS5xx
763
User Manual 1173.9628.02 ─ 16
In the Standalone (SA) scenario, this parameter is controlled by the measurement. In
the Combined Signal Path (CSP) scenario, it is controlled by the signaling application.
Remote command:
(SA)
(CSP)
4.3.2.2
Measurement Control Settings
The "Measurement Control" parameters configure the scope of the LTE PRACH mea-
surement and inform the measurement about preamble-related signal properties nor-
mally determined by higher layers.
The "Measurement Control" section consists of a general part (see below), a part with
.
While the combined signal path scenario is active, some of the measurement control
parameters display values determined by the controlling signaling application. This is
indicated in the parameter description. See also
"Scenario = Combined Signal Path"
Fig. 4-15: LTE PRACH: Measurement Control settings - general part
Repetition
Defines how often the measurement is repeated if it is not stopped explicitly or by a
failed limit check.
●
Continuous:
The measurement is continued until it is explicitly terminated; the
results are periodically updated.
●
Single-Shot:
The measurement is stopped after one statistics cycle.
Single-shot is preferable if only a single measurement result is required under fixed
conditions, which is typical for remote-controlled measurements. Continuous mode is
suitable for monitoring the evolution of the measurement results in time and observe
how they depend on the measurement configuration, which is typically done in manual
control. The reset/preset values therefore differ from each other.
Remote command: