LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
307
User Manual 1173.9628.02 ─ 16
<TX3Converter>
TX1 | TX2 | TX3 | TX4
TX module for the third output path
<TX4Connector>
RF1C | RF1O | RF2C | RF3C | RF3O | RF4C | RFAC | RFAO |
RFBC
RF connector for the fourth output path
<TX4Converter>
TX1 | TX2 | TX3 | TX4
TX module for the fourth output path
Firmware/Software:
V3.2.70
Options:
R&S CMW-KS520 and R&S CMW-KS521
Manual operation:
See
ROUTe:LTE:SIGN<i>:SCENario:CATRfout
<RXConnector>, <RXConverter>,
<PCCTXConnector>, <PCCTXConverter>, <SCCTXConnector>,
<SCCTXConverter>
Activates the scenario "2CC CA - 2 RF Out", using one RF input path, one PCC RF
output path and one SCC RF output path. Selects the RF connectors and RX/TX mod-
ules to be used. To query the active scenario, use
.
Depending on the installed hardware and the active sub-instrument or instance <i>
only a subset of the described parameter values is allowed. The *RST values and the
mapping of virtual connector names to physical connectors also depend on the active
sub-instrument or instance <i>.
All instruments are equipped with the RF 1 and RF 2 connectors and one RX and TX
module. Additional RF connectors and RX/TX modules are optional.
See also: "Signal Path Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Connector names:
●
RF1C | RF2C | RF3C | RF4C = RF 1 COM to RF 4 COM, front panel
●
RFAC | RFBC = virtual names for the RF COM connectors
●
RF1O | RF3O = RF 1 OUT | RF 3 OUT, front panel
●
RFAO = virtual name for the RF OUT connectors
Parameters:
<RXConnector>
RF1C | RF2C | RF3C | RF4C | RFAC | RFBC
RF connector for the input path
<RXConverter>
RX1 | RX2 | RX3 | RX4
RX module for the input path
<PCCTXConnector> RF1C | RF1O | RF2C | RF3C | RF3O | RF4C | RFAC | RFAO |
RFBC
RF connector for the PCC output path
Command Reference