LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
359
User Manual 1173.9628.02 ─ 16
CONFigure:LTE:SIGN<i>:UL:PUSCh:TPC:SET
................................................................ 359
CONFigure:LTE:SIGN<i>:UL:PUSCh:TPC:PEXecute
........................................................ 359
CONFigure:LTE:SIGN<i>:UL:PUSCh:TPC:RPControl
........................................................360
CONFigure:LTE:SIGN<i>:UL:PUSCh:TPC:CLTPower
....................................................... 360
CONFigure:LTE:SIGN<i>:UL:PUSCh:TPC:SINGle
............................................................ 361
CONFigure:LTE:SIGN<i>:UL:PUSCh:TPC:UDPattern
........................................................361
CONFigure:LTE:SIGN<i>:UL:PUSCh:TPC:SET
<SetType>
Selects the active TPC setup to be executed for power control of the PUSCH.
For some setups, additional commands are available for configuration or to trigger the
execution:
●
SINGle
,
UDSingle
,
RPControl
●
SINGle
●
UDSingle
,
UDContinuous
●
CLOop
●
RPControl
Parameters:
<SetType>
MINPower | MAXPower | CONStant | SINGle | UDSingle |
UDContinuous | ALT0 | CLOop | RPControl
MINPower
: command the UE to minimum power
MAXPower
: command the UE to maximum power
CONStant
: command the UE to keep the power constant
SINGle
: send a pattern once (pattern contains only one type of
TPC command)
UDSingle
: send a pattern once (mix of different TPC commands
possible)
UDContinuous
: send a pattern over and over again
ALT0
: send an alternating pattern continuously
CLOop
: command the UE to a configurable target power
RPControl
: patterns for 3GPP relative power control test
*RST:
CLO
Example:
See
Modifying Parameters for an Established Connection
Firmware/Software:
V2.0.20
V3.2.50: added RPControl
Manual operation:
See
CONFigure:LTE:SIGN<i>:UL:PUSCh:TPC:PEXecute
Execute the active TPC setup for power control of the PUSCH. This is only relevant for
setups which are not executed automatically (
SINGle
,
UDSingle
,
RPControl
).
Command Reference