![R&S CMW-KM5 Series Скачать руководство пользователя страница 791](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802791.webp)
LTE PRACH Measurement
R&S
®
CMW-KM5xx/-KS5xx
791
User Manual 1173.9628.02 ─ 16
Parameters:
<PRACHfreqOffset> Range:
0 to <total RB - 6> depending on channel band-
width, see table below
*RST:
0
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V2.0.10
Manual operation:
See
Table 4-4: Maximum input value depending on channel bandwidth
Channel Bandwidth [MHz]
1.4
3
5
10
15
20
<total RB-6>
0
9
19
44
69
94
4.5.3.4
Modulation Measurement Settings
The following commands specify settings for the modulation measurements.
CONFigure:LTE:MEAS<i>:PRACh:SCOunt:MODulation
.................................................... 791
CONFigure:LTE:MEAS<i>:PRACh:MODulation:LRSindex
..................................................791
CONFigure:LTE:MEAS<i>:PRACh:MODulation:ZCZConfig
................................................ 792
CONFigure:LTE:MEAS<i>:PRACh:MODulation:SINDex:AUTO
...........................................792
CONFigure:LTE:MEAS<i>:PRACh:MODulation:SINDex
.....................................................793
CONFigure:LTE:MEAS<i>:PRACh:MODulation:EWLength
.................................................793
CONFigure:LTE:MEAS<i>:PRACh:MODulation:EWPosition
............................................... 793
CONFigure:LTE:MEAS<i>:PRACh:SCOunt:MODulation
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<StatisticCount>
Number of measurement intervals
Range:
1 to 1000
*RST:
2
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V2.0.10
Manual operation:
See
CONFigure:LTE:MEAS<i>:PRACh:MODulation:LRSindex
<LogRootSeqIndex>
Specifies the logical root sequence index to be used for generation of the preamble
sequence.
Command Reference