LTE PRACH Measurement
R&S
®
CMW-KM5xx/-KS5xx
782
User Manual 1173.9628.02 ─ 16
The GPRF External Power Sensor measurement could not be started due to miss-
ing power sensor.
●
30 (File not Found)
:
The specified file could not be found.
●
40 (ARB File CRC Error)
:
The ARB file CRC check failed. The ARB file is corrupt and not reliable.
●
42 (ARB Header Tag Invalid)
:
The ARB file selected in the GPRF generator contains an invalid header tag.
●
43 (ARB Segment Overflow)
:
The number of segments in the multi-segment ARB file is higher than the allowed
maximum.
●
44 (ARB File not Found)
:
The selected ARB file could not be found.
●
45 (ARB Memory Overflow)
:
The ARB file length is greater than the available memory.
●
50 (Startup Error)
:
The Data Application Unit (DAU), a DAU service or a DAU measurement could not
be started. Please execute a DAU selftest.
●
51 (No Reply)
:
The DAU has received no response, for example for a ping request.
●
52 (Connection Error)
:
The DAU could not establish a connection to internal components. Please restart
the instrument.
●
53 (Configuration Error)
:
The current DAU configuration by the user is incomplete or wrong and could not be
applied. Check especially the IP address configuration.
●
54 (Filesystem Error)
:
The hard disk of the DAU is full or corrupt. Please execute a DAU selftest.
●
60 (Invalid RF-Connector Setting)
The individual segments of a list mode measurement with R&S CMWS use differ-
ent connector benches. This is not allowed. All segments must use the same
bench.
Check the "Info" dialog for the relevant segment numbers.
●
101 (Firmware Error)
:
Indicates a firmware or software error. If you encounter this error for the first time,
restart the instrument.
If the error occurs again, consider the following hints:
–
Firmware errors can often be repaired by restoring the factory default settings.
To restore these settings, restart your instrument and press the "Factory
Default" softkey during startup.
–
If a software package (update) has not been properly installed this is often indi-
cated in the "Setup" dialog, section "SW/HW-Equipment > Installed Software".
–
A software update correcting the error may be available. Updates are e.g. pro-
vided in the "CMW Customer Web" on GLORIS (registration required):
.