LTE SRS Measurement
R&S
®
CMW-KM5xx/-KS5xx
821
User Manual 1173.9628.02 ─ 16
Statistical Results
The statistical values in the table are calculated as follows:
●
Current:
Value of the result obtained in the last measurement interval.
●
Average:
Average of all "Current" values referenced to the last statistics cycle.
●
Min, Max:
Largest or smallest "Current" value that the R&S
CMW obtained since
the start of the measurement.
●
StdDev:
Standard deviation of all "Current" values since the start of the measure-
ment.
All statistical results (statistical tables and "Average" or "Max" traces) are calculated
according to the general rules for statistical results.
See also: "Statistical Results" in the R&S
CMW user manual, chapter "System Over-
view"
Statistic Count
Progress bar for the measurement. During the first single shot after the start of the
measurement, the bar shows the number of completed measurement intervals relative
to the "Statistic Count". A filled progress bar indicates that the first shot is complete
and the statistical depth has been reached.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "System Over-
view"
Out of Tolerance
Percentage of measurement intervals (SRS symbols) that were failed because they
exceeded the limits in the diagram.
5.2.6.1
Modifying Views
Use the "Display" parameters to change the appearance and contents of the view. The
following "Display" hotkeys are available at the bottom of the GUI:
Hotkey
Description
"Select Trace ..."
Select the trace types to be displayed in the view.
"X Scale... / Y Scale..."
Modify the ranges of the X-axis and the Y-axis.
5.2.6.2
Using Markers
Use the "Marker" parameters to activate markers and to modify their position. The fol-
lowing "Marker" hotkeys are available at the bottom of the GUI: