![R&S CMW-KM5 Series Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802078.webp)
LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
78
User Manual 1173.9628.02 ─ 16
nario in the measurement. As a result the measurement application uses the most
important settings of the signaling application, e.g. the RF settings.
Proceed as follows:
1. Use the "LTE TX Meas" softkey to switch to the multi evaluation measurement.
The measurement application is opened and the combined signal path scenario is
selected automatically. Furthermore the frame trigger signal provided by the "LTE
Signaling" application is selected as trigger source.
2. Press "ON | OFF" to start the measurement.
The main view provides an overview of the measurement results.
To enlarge a diagram perform one of the following actions:
●
Double-click it using a connected mouse.
●
Select it by turning the rotary knob and open it by pressing the rotary knob.
2.3.1.4
Possible Extensions
While the connection is established, you can vary "LTE Signaling" settings and
observe the behavior of the UE under test using the "LTE Multi Evaluation" measure-
ment. E.g. you can configure the signaling application to send Transmit Power Control