FETCh:LTE:SIGN<i>:EBLer:TRACe:THRoughput[:PCC]?
FETCh:LTE:SIGN<i>:EBLer:TRACe:THRoughput:SCC?
Returns the throughput trace for the sum of all PCC DL streams or all SCC DL
streams.
Each value is returned as a pair of X-value and Y-value. The number of result pairs n
equals the number of subframes to be processed per measurement cycle, divided by
200.
Returned results: <Reliability>, <XValue>
1
, <YValue>
1
, ..., <XValue>
n
, <YValue>
n
Return values:
<Reliability>
<XValue>
Subframe label, 0 = last processed subframe, -1 = previously
processed subframe and so on
Range:
-199800 to 0
<YValue>
Throughput value calculated from the BLER result of 200 pro-
cessed subframes (the labeled subframe and the previous 199
subframes)
Range:
0 kbit/s to 200E+3 kbit/s
Default unit: kbit/s
Example:
See
Performing a Single-Shot BLER Measurement
Usage:
Query only
Firmware/Software:
V2.0.10
V3.2.50: SCC command added
Options:
R&S CMW-KS502/-KS552 (FDD/TDD) for SCC
FETCh:LTE:SIGN<i>:EBLer:TRACe:THRoughput:ALL?
Returns the throughput trace for the sum of all downlink streams and all carriers (PCC
plus SCC).
Each value is returned as a pair of X-value and Y-value. The number of result pairs n
equals the number of subframes to be processed per measurement cycle, divided by
200.
Returned results: <Reliability>, <XValue>
1
, <YValue>
1
, ..., <XValue>
n
, <YValue>
n
Return values: