LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
557
User Manual 1173.9628.02 ─ 16
The "Lagging" time interval is excluded at the end of each subframe; if SRS signals
are allowed, at the end of each shortened subframe.
Remote command:
3.3.5.4
Spectrum Settings
The following "Measurement Control" parameters configure the spectrum measure-
ment settings.
Fig. 3-42: Measurement control settings - spectrum
Emission Mask / ACLR > Statistic Count
.................................................................... 557
...................................................................................... 557
Emission Mask / ACLR > Statistic Count
Defines the number of measurement intervals per measurement cycle for emission
mask and ACLR measurements.
One measurement interval is completed when the R&S
CMW has measured a full slot
of the configured "Measure Subframe".
Remote command:
Emission Mask > Meas Filter
Selects the resolution filter type for filter bandwidths of 100 kHz and 1 MHz (gaussian
or bandpass filter). For 30 kHz filters the type is fixed (gaussian). The filter bandwidth
to be used can be configured per emission mask area as part of the limits.
Remote command:
GUI Reference