LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
113
User Manual 1173.9628.02 ─ 16
The state "RDY" means that the signaling application is ready to receive an inter-RAT
handover from another signaling application (e.g. from WCDMA). This state is initiated
by the application acting as source of the handover.
Remote command:
Connection control hotkeys
Any interaction with a UE requires an LTE downlink signal (cell). As soon as the signal
is available (state ON, no hour glass), connection control hotkeys appear in the hotkey
bar. In data application mode, the "Connect" hotkey is even available without LTE
downlink signal, so that you can preconfigure the setup of dedicated bearers.
The available hotkeys depend on the current packet switched state, RRC state, SCC
state and connection type. For SCC activation there are several possible modes, influ-
encing the available SCC hotkeys, see
The possible hotkeys are listed in the following table.
Hotkey
Description
"Connect"
See
"Disconnect"
Release a dedicated PCC connection in test mode.
"Detach"
Send a detach request to the UE and detach the UE (independent of UE reac-
tion to the request).
"Send SMS"
Send an SMS message to the UE.
"Inter/Intra-RAT"
See
"SCC On"
Switch on the SCC DL signal.
"SCC Off"
Switch off the SCC DL signal.
"SCC add RRC"
"SCC delete RRC"
Add/delete SCC RRC connection.
"SCC activate MAC"
"SCC deactivate MAC"
Activate/deactivate MAC for the SCC.
Remote command:
Connect (hotkey)
The behavior of the "Connect" hotkey depends on the parameter
.
For connection type "Testmode", you must switch on the cell signal to make the hotkey
available. The hotkey sets up a single dedicated bearer for the PCC.
For connection type "Data Application", the hotkey is even available if the cell signal is
switched off. The hotkey opens a dialog box for configuration, connection and release
of multiple dedicated bearers. Option R&S CMW-KS510 is required.