LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
310
User Manual 1173.9628.02 ─ 16
All instruments are equipped with the RF 1 and RF 2 connectors and one RX and TX
module. Additional RF connectors and RX/TX modules are optional.
See also: "Signal Path Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Connector names:
●
RF1C | RF2C | RF3C | RF4C | = RF 1 COM to RF 4 COM, front panel
●
RFAC | RFBC = virtual names for the RF COM connectors
●
RF1O | RF3O = RF 1 OUT | RF 3 OUT, front panel
●
RFAO = virtual name for the RF OUT connectors
●
IQ2O | IQ4O | IQ6O | IQ8O = DIG IQ OUT 2 to 8, rear panel
Parameters:
<RXConnector>
RF1C | RF2C | RF3C | RF4C | RFAC | RFBC
RF connector for the input path
<RXConverter>
RX1 | RX2 | RX3 | RX4
RX module for the input path
<TXConnector>
RF1C | RF1O | RF2C | RF3C | RF3O | RF4C | RFAC | RFAO |
RFBC
RF connector for the output path
<TXConverter>
TX1 | TX2 | TX3 | TX4
TX module for the output path
<IQConnector>
IQ2O | IQ4O | IQ6O | IQ8O
DIG IQ OUT connector for external fading of the output path
Example:
See
Firmware/Software:
V3.0.10
Options:
R&S CMW-KS510
Manual operation:
See
ROUTe:LTE:SIGN<i>:SCENario:SCFading:INTernal
<RXConnector>,
<RXConverter>, <TXConnector>, <TXConverter>, <Fader>
Activates the scenario "1 Cell - Fading - 1 RF Out" with internal fading, selects the RF
input (RX) and output (TX) path (RF connectors and RX/TX modules) and the I/Q
board to be used for fading.
To query the active scenario, use
.
Depending on the installed hardware and the active sub-instrument or instance <i>
only a subset of the described parameter values is allowed. The *RST values and the
mapping of virtual connector names to physical connectors also depend on the active
sub-instrument or instance <i>.
All instruments are equipped with the RF 1 and RF 2 connectors and one RX and TX
module. Additional RF connectors and RX/TX modules are optional.
Command Reference