LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
726
User Manual 1173.9628.02 ─ 16
Firmware/Software:
V2.1.30
Options:
R&S CMW-KM012
3.5.4 Combined Signal Path Commands
For some settings, the command to be used depends on the active scenario. While the
Combined Signal Path (CSP) scenario is active, these settings are configured via com-
mands of the signaling application. While the Standalone (SA) scenario is active, they
are configured via measurement commands.
The following mapping tables provide an overview for general measurement settings
and for multi evaluation measurement commands.
Table 3-2: Mapping for general measurement settings
Setting
Commands for SA scenario
Commands for CSP scenario
Duplex mode
Connector, converter
ROUTe:LTE:SIGN<i>:SCENario:...
External attenuation
Band
Frequency, channel
Frequency offset
Expected nominal
power
User margin
Mixer level offset
Channel bandwidth
Command Reference