![R&S CMW-KM5 Series Скачать руководство пользователя страница 835](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802835.webp)
LTE SRS Measurement
R&S
®
CMW-KM5xx/-KS5xx
835
User Manual 1173.9628.02 ─ 16
5.4.5 Specifying Limits
// *****************************************************************************
// Define power dynamics limits.
// *****************************************************************************
CONFigure:LTE:MEAS:SRS:LIMit:PDYNamics ON,5.1,-10.3,-48.8
5.4.6 Performing Single-Shot Measurements
// *****************************************************************************
// Start single-shot measurement, return average power dynamics trace.
// Query the measurement state (should be "RDY").
// *****************************************************************************
READ:LTE:MEAS:SRS:TRACe:PDYNamics:AVERage?
FETCh:LTE:MEAS:SRS:STATe?
// *****************************************************************************
// Read other traces obtained in the last
// measurement without re-starting the measurement.
// *****************************************************************************
FETCh:LTE:MEAS:SRS:TRACe:PDYNamics:CURRent?
FETCh:LTE:MEAS:SRS:TRACe:PDYNamics:MAXimum?
// *****************************************************************************
// Read statistical results obtained in the last measurement
// without re-starting the measurement.
// *****************************************************************************
FETCh:LTE:MEAS:SRS:PDYNamics:CURRent?
FETCh:LTE:MEAS:SRS:PDYNamics:AVERage?
FETCh:LTE:MEAS:SRS:PDYNamics:MINimum?
FETCh:LTE:MEAS:SRS:PDYNamics:MAXimum?
FETCh:LTE:MEAS:SRS:PDYNamics:SDEViation?
// *****************************************************************************
// Read limit check results obtained in the last measurement
// without re-starting the measurement.
// *****************************************************************************
CALCulate:LTE:MEAS:SRS:PDYNamics:CURRent?
CALCulate:LTE:MEAS:SRS:PDYNamics:AVERage?
CALCulate:LTE:MEAS:SRS:PDYNamics:MINimum?
CALCulate:LTE:MEAS:SRS:PDYNamics:MAXimum?
CALCulate:LTE:MEAS:SRS:PDYNamics:SDEViation?
5.4.7 Single-Shot and Continuous Measurements
// *****************************************************************************
// Start single-shot measurement, return current power dynamics trace,
// return average trace (without repeating the measurement).
Programming
深圳德标仪器
135-1095-0799