![R&S CMW-KM5 Series Скачать руководство пользователя страница 660](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802660.webp)
LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
660
User Manual 1173.9628.02 ─ 16
FETCh:LTE:MEAS<i>:MEValuation:TRACe:IQ:LOW?
FETCh:LTE:MEAS<i>:MEValuation:TRACe:IQ:HIGH?
Returns the results in the I/Q constellation diagram for low and high EVM window posi-
tion.
See also
chapter 3.2.6.7, "Detailed Views: I/Q Constellation Diagram"
Return values:
<Reliability>
<I_Phase_1>
<Q_Phase_1> ...
<I_Phase_n>
<Q_Phase_n>
n normalized I and Q amplitudes, one value per modulation sym-
bol
Range:
-2 to 2
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V1.0.10.1
3.5.3.22
Power Dynamics Results (Traces)
The following commands return the results in the power dynamics diagram.
FETCh:LTE:MEAS<i>:MEValuation:TRACe:PDYNamics:CURRent?
FETCh:LTE:MEAS<i>:MEValuation:TRACe:PDYNamics:AVERage?
FETCh:LTE:MEAS<i>:MEValuation:TRACe:PDYNamics:MAXimum?
READ:LTE:MEAS<i>:MEValuation:TRACe:PDYNamics:CURRent?
READ:LTE:MEAS<i>:MEValuation:TRACe:PDYNamics:AVERage?
READ:LTE:MEAS<i>:MEValuation:TRACe:PDYNamics:MAXimum?
Return the values of the power dynamics traces. Each value is sampled with 48 T
s
,
corresponding to 1.5625 µs. The results of the current, average and maximum traces
can be retrieved.
See also
chapter 3.2.6.10, "Detailed Views: Power Dynamics"
Return values:
<Reliability>
<Power>
2048 power values, from -1100 µs to +2098.4375 µs relative to
the start of the measure subframe. The values have a spacing of
1.5625 µs. The 705
th
value is located at the start of the "Mea-
sure Subframe" (0 µs).
The diagram at the display shows only a subsection of this trace,
depending on the selected time mask.
Range:
-100 dBm to 55 dBm
Default unit: dBm
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V2.0.10
Command Reference
深圳德标仪器