LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
82
User Manual 1173.9628.02 ─ 16
5. Press OK to close the configuration dialog.
6. Press ON | OFF.
The application starts and the "Iperf" softkey indicates the "RUN" state.
The R&S
CMW sends data packets to the UE, using the downlink bit rate indicated
in the last column and the graphic.
7. Configure the IPerf tool at UE side so that it listens to the UDP port configured for
the used client entry.
8. Start the IPerf tool at UE side and compare the bit rates received at the UE with the
sent bit rates at the R&S
CMW.