![R&S CMW-KM5 Series Скачать руководство пользователя страница 609](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802609.webp)
LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
609
User Manual 1173.9628.02 ─ 16
Parameters:
<PhsLayerCellID>
Range:
0 to 503
*RST:
0
Example:
See
Firmware/Software:
V1.0.10.1
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:DSSPusch
<DeltaSeqShift>
Specifies the delta sequence shift value (Δ
ss
) used to calculate the sequence shift pat-
tern for PUSCH.
Parameters:
<DeltaSeqShift>
Range:
0 to 29
*RST:
0
Example:
See
Firmware/Software:
V1.0.10.1
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:GHOPping
<Value>
Specifies whether group hopping is used or not.
This command is only relevant for the standalone scenario. For the combined signal
path scenario, use
Parameters:
<Value>
OFF | ON
*RST:
OFF
Example:
See
Firmware/Software:
V2.0.20
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:MSUBframes
<SubframeOffset>,
<SubframeCount>, <MeasSubframe>
Configures the scope of the measurement, i.e. which subframes are measured.
Parameters:
<SubframeOffset>
Start of the measured subframe range relative to the trigger
event
Range:
0 to 9
*RST:
0
Command Reference