LTE SRS Measurement
R&S
®
CMW-KM5xx/-KS5xx
827
User Manual 1173.9628.02 ─ 16
5.3.2.2
Measurement Control Settings
The "Measurement Control" parameters configure the scope of the LTE SRS measure-
ment.
While the combined signal path scenario is active, some of the measurement control
parameters display values determined by the controlling signaling application. This is
indicated in the parameter description. See also
"Scenario = Combined Signal Path"
Fig. 5-7: LTE SRS: Measurement Control settings
Repetition
Defines how often the measurement is repeated if it is not stopped explicitly or by a
failed limit check.
●
Continuous:
The measurement is continued until it is explicitly terminated; the
results are periodically updated.
●
Single-Shot:
The measurement is stopped after one statistics cycle.
Single-shot is preferable if only a single measurement result is required under fixed
conditions, which is typical for remote-controlled measurements. Continuous mode is
suitable for monitoring the evolution of the measurement results in time and observe
how they depend on the measurement configuration, which is typically done in manual
control. The reset/preset values therefore differ from each other.
Remote command:
Stop Condition
Specifies the conditions for an early termination of the measurement:
●
None:
The measurement is performed according to its "Repetition" mode and "Sta-
tistic Count", irrespective of the limit check results.
●
On Limit Failure:
The measurement is stopped as soon as one of the limits is
exceeded, irrespective of the repetition mode set. If no limit failure occurs, it is per-
formed according to its "Repetition" mode and "Statistic Count". Use this setting for
measurements that are essentially intended for checking limits, e.g. production
tests.
Remote command:
GUI Reference