LTE PRACH Measurement
R&S
®
CMW-KM5xx/-KS5xx
793
User Manual 1173.9628.02 ─ 16
CONFigure:LTE:MEAS<i>:PRACh:MODulation:SINDex
<SequenceIndex>
Specifies the sequence index, i.e. which of the 64 preamble sequences of the cell is
used by the UE. This setting is only relevant if automatic detection is disabled, see
.
Parameters:
<SequenceIndex>
Range:
0 to 63
*RST:
0
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V2.0.10
Manual operation:
See
CONFigure:LTE:MEAS<i>:PRACh:MODulation:EWLength
<LengthFormat0>,
<LengthFormat1>, <LengthFormat2>, <LengthFormat3>, <LengthFormat4>
Specifies the EVM window length in samples for all preamble formats.
Parameters:
<LengthFormat0>
Length for preamble format 0
Range:
1 to 3168
*RST:
3072
<LengthFormat1>
Length for preamble format 1
Range:
1 to 21024
*RST:
20928
<LengthFormat2>
Length for preamble format 2
Range:
1 to 6240
*RST:
6144
<LengthFormat3>
Length for preamble format 3
Range:
1 to 21024
*RST:
20928
<LengthFormat4>
Length for preamble format 4
Range:
1 to 448
*RST:
432
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V2.0.10
V2.0.20: *RST values modified
Manual operation:
See
CONFigure:LTE:MEAS<i>:PRACh:MODulation:EWPosition
<EVMwindowPos>
Specifies the position of the EVM window used for calculation of the trace results.
Command Reference