Copyright 2022 Baker Hughes Company.
196 | DPI610E Instruction Manual–English
Chapter 11. Documenting
Step 3
Pressure Sensor Function
For pressure-to-pressure calibrations i.e.,
INT
pressure and
EXT
pressure, the option
to swap the assignments of the sensors as
Input and Output using the
Toggle
softkey.
Care must be taken to make sure the sensor
type and range are compatible with that
specified in the test.
Step 4
10 V/24 V Power
The option to enable the internal 10 V or
24 V power is also available in the
Calibrator Tasks chapter (see Section 6.2.9
on page 88 menu). Select the required
option from the settings screen.
The tick-box is ticked when enabled and
unticked when disabled.