English–DPI610E Instruction Manual | 259
EXTERNAL RTD SENSOR CALIBRATION MENU
Step 4
Press the
Next
softkey to complete
Calibration Point 1
and proceed to
Calibration Point 2
.
Step 5
Repeat Steps 2 to 4, but this time begin
by applying the nominal positive full-
scale resistance or temperature value to
the DPI 610E. This completes
Calibration Point 2
.
Step 6
After pressing
Next
from the previous
step, a message window will appear,
requesting confirmation to perform the
calibration adjustment based on the
calibration points used in the procedure.
Select
OK
to apply the calibration
adjustment.
Otherwise, select the
Cancel
button to
return to the
Calibration
screen