11. Documenting
11.1 Overview
Use the Documenting application to do documented calibration of Device Under Test (DUT)
equipment or of assets using defined test procedures.
Test procedures can be created within the Documenting application and stored for future use.
When test procedures are used to calibrate DUTs, the test details and results of the calibration
are saved in DPI 610E memory. This data can also be transferred to a PC for further analysis or
post calibration tasks.
A calibration certificate template wizard is provided by Druck to transform the data into a
professional looking format for printing or filing purposes.
11.1.1 How to start the Documenting application
11.2 INTERNAL PROCEDURES
11.2.1 How to select the INTERNAL PROCEDURES mode
Only the
INTERNAL PROCEDURES
mode is available in the current software release. There will
be a
REMOTE PROCEDURES
mode in the future.
Select the
Documenting
softkey from the
Dashboard. This will show the
Procedures
screen.
Step 1
To view, create or run internal
procedures, select the
INTERNAL
PROCEDURES
softkey from the
Procedures
screen. This will show a
list of available internal procedure files.
Documenting