106 | DPI610E Instruction Manual–English
Chapter 6. Calibrator Tasks
6.2.11 Millivolts Measure - Configuration
Voltage can also be measured in millivolts (mV) and is a
function of
CH2
only. When the
Millivolts
function is
selected, the
Direction
is automatically set to
Measure
as the Source option is not supported by the DPI 610E.
The supported millivolts range is -2000 mV to 2000 mV.
When using the
Millivolts
function, there is the
additional option to use the internal 10 V (non-IS units
only) or 24 V power supply provided by DPI 610E or to
measure the millivolts input from a source that is
powered external to the unit.
Terminal connections for measurement of millivolts are
shown in diagram on the left.
Step 1
From the
Calibrator
screen, select the
Channel 2
Setup
(either press the
screen icon or press the softkey.)
Step 2
Select
FUNCTION
from the
CH2
setup
screen.