260 | DPI610E Instruction Manual–English
Chapter 13. Instrument Calibration
Step 7
If
OK
is selected, the
CALIBRATION
COMPLETE
message appears in the
status box, to confirm the adjustment
has been made.
At this stage, there are three icons for
three ways to proceed. These are:
VERIFICATION
softkey navigates
to the Verification screen (see Step 8).
RESTART
softkey allows the
calibration procedure to be restarted if a
repeat calibration is required.
BACK
softkey exits the calibration
procedure and returns the user to the
Perform Calibration
menu screen.
Step 8
The verification mode shows live
reference and measured (or sourced)
values.
Here, different values or points within
the measured (or sourced) range can be
checked to verify the adjustment is
satisfactory.
Once verification is complete, select the
Back
icon to exit the calibration
procedure or select the
Restart
softkey to repeat the calibration.