Chapter 12. HART
Step 9
Select the
Verification
softkey to test if
the adjustment has been sufficient.
Step 10
In the
Verification
screen, use the
Edit
softkey to enter test current output
values between 0 and 24 mA and check
the measured values at each point.
To exit the Verification screen, use the
Back
softkey.