Copyright 2021 Baker Hughes Company.
120 | DPI610E Instruction Manual–English
Chapter 6. Calibrator Tasks
Step 3
Tap the
ALARM
area or use the
Navigation Pad buttons to select it. The
Alarm
screen appears.
Step 4
Use the procedures given below to set
the
LOW
and
HIGH
values for alarm
activation.
Use the Navigation button to move to
the
LOW
option, press the Navigation
Enter
button to show an onscreen
keypad. Use the keypad to enter the
value for the lower end of normal range
condition. Select the
Tick
softkey to
confirm the value.
Note
: By default, the
LOW
(zero or
negative full-scale) range value of the
selected measure function is populated.
Select the
HIGH
option and enter the
value for the highest end of normal
range condition. Select the
Tick
softkey
to confirm the value.
Note
: by default, the
HIGH
(positive full-
scale) range value of the selected
measure function is populated.
Use the Navigation Pad to select the
ALARM
row. Press the
Enter
key in the
pad to insert a tick into its checkbox or
tap in the checkbox. Press the Tick
softkey to confirm the alarm settings
and return to the
PROCESS
options
screen.
Note
: To disable the Alarm, select the
Alarm Checkbox to remove the
Tick
mark.