English–DPI 610E Instruction Manual | 31
Pneumatic System
2.1.7 Generating Pressure/Vacuum
Step 1
Seal the system by turning the pressure release valve fully clockwise.
CAUTION
Do not submerge the instrument in the pressure medium. If fluid
enters any part of the electrical connection panel, allow the instrument to dry out
thoroughly before use.
Step 2
Adjust the pressure-vacuum selector to pressure mode (+) or vacuum
mode (-) depending on the desired operation.
Note:
To adjust the selector from pressure mode to vacuum mode,
turn it clockwise. To adjust the selector from vacuum mode to
pressure mode, turn it counter-clockwise.
CAUTION
To prevent the instrument from sliding, make sure that either the
slip prevention feet are engaged or the instrument is secured against a flat
surface.
Step 3
Use the pump to achieve the desired pressure/vacuum.
Step 4
Turn the volume adjuster to make fine adjustments to the
pressure/vacuum.