EXTERNAL PRESSURE SENSOR CALIBRATION MENU
13.8.3 SET CALIBRATION DATE & INTERVAL
The
External Pressure Sensor Calibration Date & Interval
screen provides the following
options:
.
Figure 13-5: External Sensor Calibration Date & Interval Screen
To access the
SET CALIBRATION DATE & INTERVAL
screen:
Option
Description
LAST CALIBRATION
Set the date the instrument was last calibrated
CALIBRATION INTERVAL
Enable user notification by setting the number of
days between the last calibration and the next
scheduled calibration (default is 365 days)
CALIBRATION DUE
Enable user notification by setting a specific date to
schedule next calibration (default is based on the
date of the last calibration and the specified
calibration interval)
Step 1
Select
SET CALIBRATION DATE &
INTERVAL
screen
from the
External
Sensor
Calibration
screen.