![Druck DPI 610E Instruction Manual Download Page 145](http://html1.mh-extra.com/html/druck/dpi-610e/dpi-610e_instruction-manual_2533086145.webp)
Copyright 2022 Baker Hughes Company.
146 | DPI610E Instruction Manual–English
Chapter 8. External Sensors
If a channel is configured with the
External Pressure
function without a sensor connected, the
primary reading in that channel window will be appear as “
- - - - -
“.
If the external sensor is disconnected while the function is in use an error message may appear.
PM 700E External Remote sensors with DK481 firmware version 2.00.00 and above are fully
supported for use with the DPI 610E. Sensor validation is done each time a connection is
detected
If a channel is configured for External Pressure and an unsupported sensor is detected (for
example a PM 700E sensor with firmware older than 2.00.00), a warning message appears
stating the sensor is not supported and the status bar will have the sensor icon
in it..
The readings from the sensor will appear as normal when configured on the Calibrator screen
however, sensor calibration will not be supported using the DPI 610E. All PM700E sensors
etched with a part number ending in “-3” have the correct firmware installed.
If a channel is configured for External Pressure and an unrecognised sensor is detected, a
warning message appears stating the connected sensor is not compatible and the sensor icon
will appear in the status bar.