Chapter 13. Instrument Calibration
Step 10
Once again, apply Steps 6 to 8, but this time
begin by applying the positive full-scale
Current (approximately 20 mA) to the
DPI 610E.
Press the
Next
softkey to complete
Calibration point 3.
Step 11
A message window appears, requesting
confirmation to perform the calibration
adjustment based on the calibration points
used in the procedure.
Select
OK
to apply the calibration
adjustment.
Otherwise, select the
Cancel
button to
return to the
Calibration
screen.
Step 12
If
OK
is selected, the
CALIBRATION
COMPLETE
message appears in the status
box, to confirm the adjustment has been
made.
At this stage, there are three icons for three
ways to proceed. These are:
VERIFICATION
softkey navigates to
the Verification screen (see Step 13).
RESTART
softkey allows the
calibration procedure to be restarted if a
repeat calibration is required.
BACK
softkey exits the calibration
procedure and returns the user to the
Perform Calibration
menu screen.