Documenting Settings
11.4 Documenting Settings
Step 1
If additional test procedure setup is required
before documenting begins, select the
Setup
softkey.
Step 2
Pressure Sensor Type
For pressure calibrations, if a different
pressure sensor type is to be used from that
which was used to create the test
procedure, it is possible to change the
sensor configuration to match.
Note
: This applies to the
INT
and
EXT
Pressure functions used as the Input or/and
Output only.
For example, if a gauge pressure sensor is
specified in the test procedure and the
available pressure sensor is an absolute
pressure sensor, the absolute pressure
sensor can be configured to pseudo gauge
(pneumatic variants only) or sealed gauge
(hydraulic variants only).
To use
PSEUDO RANGE
, select the
relevant pseudo-range option from the
settings screen. The tick-box is ticked when
enabled and unticked when disabled.