![Druck DPI 610E Instruction Manual Download Page 95](http://html1.mh-extra.com/html/druck/dpi-610e/dpi-610e_instruction-manual_2533086095.webp)
Copyright 2021 Baker Hughes Company.
96 | DPI610E Instruction Manual–English
Chapter 6. Calibrator Tasks
1.
NUDGE
This is the default
Source Automation
option. It allows the source value to be incremented or
decremented by a defined step size value. This can be achieved in the current source channel
window by:
Step 1
To use the screen
.
From the
Source Automation
screen
select the
NUDGE
radio
button and then
the
Edit
softkey to show the
Nudge
Automation
screen.
The Nudge Automation data will
appear. Go to step 2 to change this data
if required.
Click the
Proceed
icon to activate
the option and return to the
Calibrate
screen.
To use the Navigation Pad
.
When using the keypad, press the
DOWN
button to select the
CH2
window. Press the
Edit
softkey to
show the
Nudge Automation
screen.
The Nudge Automation data will
appear. Go to step 2 to change this data
if required.
Step 2
Select the
STEP SIZE
row to show an
onscreen keypad, enter the new value
using the onscreen or Navigator keys.
Tap the
Back
icon to enter the
value and return to the
Calibrator
screen.
For a description of
Scaling
refer to