Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual | 241
PERFORM CALIBRATION
Step 5
Press the
Next
softkey to complete
Calibration point 1
and proceed to
Calibration point 2
.
Step 6
Apply Steps 3 to 5, but this time begin by
applying the nominal positive full-scale
pressure to the DPI 610E.
When the measured Pressure reading is
stable, select the
Next
icon to continue.
This completes
Calibration point 2
.
Step 7
After pressing
Next
from the previous step,
a message window will appear, requesting
confirmation to perform the calibration
adjustment based on the calibration points
used in the procedure.
Select
OK
to apply the calibration
adjustment.
Otherwise, select the
Cancel
button to
return to the
Calibration
screen