154 | DPI610E Instruction Manual–English
Chapter 8. External Sensors
Tap on the desired profile to select it. Tap on the
View
icon softkey to view or/and edit the
coefficients and parameters within the profile. Alternatively, use the
UP
/
DOWN
Navigation Pad
buttons to select the required profile and press the
View
softkey to view or/and edit the
coefficients and parameters within the profile.
To edit any of the parameters:
•
Tap on the value field of the desired parameter:
•
Use the
UP
/
DOWN
buttons to move the selector icon to the required parameter to
change. Press the
Enter
button to enter the edit screen of the parameter. Press or tap
the
Tick
icon softkey to accept and save the new values.
It is possible to copy the parameters saved in any of the profiles to a custom profile. Note the
Default EU and Default US profiles cannot be overwritten but can be copied into any of the
custom profiles. Custom profiles are saved in the RTD-INTERFACE not on the DPI 610E.
When all parameter fields are filled, use the
Back
softkey to go back to the
RTD Profile
screen.
Table 8-5: RTD Profile parameters
Parameter
Description
Identifier
Give the custom profile a unique label
RTD Type
Select from 2, 3, 4 wire RTD configuration. 4-wire is default.
Coefficient Format
Select from Callendar-Van Dusen equation formats or use the ITS90
equation format.
Coefficient R0
Resistance at 0°C
Coefficient 1 (a, A, Alpha)
First coefficient value
Coefficient 2 (b, B, Beta)
Second coefficient value
Coefficient 3 (c, C, Delta)
Third coefficient value
Min. Temperature
Minimum temperature
Max. Temperature
Maximum temperature