Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual | 237
PERFORM CALIBRATION
Step 6
Press the
Next
softkey to complete
Calibration point 1
and proceed to
Calibration point 2
.
Step 7
Apply Steps 4 to 6, but this time begin by
applying the nominal positive full-scale
pressure to the DPI610E.
Press the
Next
softkey to complete
Calibration point 2
and proceed to
Calibration point 3
.
Step 8
A message window appears, requesting
confirmation to perform the calibration
adjustment based on the calibration points
used in the procedure.
Select
OK
to apply the calibration
adjustment.
Otherwise, select the
Cancel
button to
return to the
Calibration
screen.